Measuring capacitors and inductivities not related to speakers

redmountain

New Member
Thread Starter
Joined
Dec 8, 2022
Posts
31
Is this possible,
and how exactly ?
Sense resistor values ?
YT videos ?
Thank youand best regards

Klaus
 

luigi57

Registered
Joined
Nov 27, 2024
Posts
6
Hi John,
I'm using your fantastic software to test audio transformer for MC cartridge; my task is to measure all transformer parameters to simulate it.The transformer I want to test is an old Audio Tecnica AT660T with a step-up ratio ranging from 1:34 to 1:125: do you suggest to measure impedance from the secondary side in order to have a step-down transformer?
Why low-frequency is limited to 2Hz? Is there any possibility to extend to 0.2Hz?
Many thanks for your suggestions.
Luigi
 

John Mulcahy

REW Author
Joined
Apr 3, 2017
Posts
8,115
You can change the graph start frequency in the View preferences. I don't have any advice on transformer measurement, sorry.
 

dcibel

Member
Joined
Sep 10, 2017
Posts
293
Sweep can start at 0Hz, but IR window needs to be >500ms to capture frequencies below 2Hz.To capture frequencies to 0.2Hz requires a window of 5 seconds.
 

luigi57

Registered
Joined
Nov 27, 2024
Posts
6
Thank you Decibel.
Do you mean that to capture frequencies to 0.2Hz only the right IR Window must be >5 seconds for sine sweep, or even the left IR Window must be changed?
 

dcibel

Member
Joined
Sep 10, 2017
Posts
293
0.2Hz wavelength is 5 seconds long, so total window needs to be at least 5 seconds, simple as that. How much left and right depends on where the reference marker is located. If ref at IR start, the 5s right, 0 left as bare minimum
 

dcibel

Member
Joined
Sep 10, 2017
Posts
293
Keep in mind that most audio interfaces have DC blocking caps on the input, so you may not find very high signal level at 0.2Hz, SNR could present some problems in accuracy/reliability of results. As well, both channels of audio input need to be well matched, which will primarily be affected by component tolerances in that DC blocking high pass circuit. Simple loopback test of each input channel will confirm on your equipment.
 

luigi57

Registered
Joined
Nov 27, 2024
Posts
6
My sound card has -3dB points at 1,4Hz and 80KHz and the interface to sound card has -3dB points at 0,16Hz and 250KHz. My interest is on measuring impedance and amplifier, not speakers response. SNR, for sure, can limit low-level results.
 

trobbins

Member
VIP Supporter
Joined
Nov 26, 2017
Posts
133
Location
Australia
-3dB at 1.4Hz for a loopback is a good start for using the low frequency capability of REW.

My EMU 0404 has a DAC/headphone output with -3dB at 1.5Hz (up to 94kHz), but the ADC/input amp coupling is far more limiting with -3dB at 2.8Hz at 44.1/48kHz sampling, rising to 5Hz at 192kHz sampling (even though the AK5385 HPF -3dB spec is down at 1Hz for 44/48kHz sampling). With respect to using a compensation curve to flatten the loopback response, the compensation is -11dB and 140 deg phase at 2Hz, so some heavy lifting from the compensation when measuring below 2Hz. The bulk of that low frequency roll-off is from 47uF 63V coupling caps, which can be swapped out for higher capacitance parts (and even at lower voltage if phantom power is disabled), although it could be practical but quite risky to bypass them for dc response if the soundcard was dedicated to a particular testing setup.
 

redmountain

New Member
Thread Starter
Joined
Dec 8, 2022
Posts
31
I set up impedance measurement
Sense resistor is 66 Ohms
Will this work for up to 4 Hy coils / 68 Ohm resistance ?
Cap up to 3 µ F read ok

Klaus
 

trobbins

Member
VIP Supporter
Joined
Nov 26, 2017
Posts
133
Location
Australia
Hi Klaus, it looks like you may be measuring iron or ferrite cored inductors? If so then their inductance may be significantly affected applied signal voltage level, which may then need to be part of the measurement reporting. Does your application also pass dc through the inductor?
 

redmountain

New Member
Thread Starter
Joined
Dec 8, 2022
Posts
31
Yes
> https://ferrite.de/en/products/p-cores/p18-slash-11/
I developped with these some 40 years ago
octave LC filter EQs with +-20dB and FET circuit...
And now maybe a revival...This thing sounds terrific :-))
I am Measuring with a Mac headphone out @ about 5dB headroom FS
Should I compare @ other levels ?
I struggled with the 2 half ferrites by applying different force on
holding them together...
...Measuring with REW the inductance

Thanks

Klaus
 

trobbins

Member
VIP Supporter
Joined
Nov 26, 2017
Posts
133
Location
Australia
Ta - so a P18 pot core - and yes very common 40 years ago 👍.

Signal voltage shouldn't then be a concern wrt the core material. The measured inductance may have a large tolerance between samples if the coreset is ungapped. Do you only have ungapped coresets, or some gapped sets? Are you using the default core clamping clips? Do the coresets include centre leg adjustment?
 

redmountain

New Member
Thread Starter
Joined
Dec 8, 2022
Posts
31
@ John
REWv 5.31.3
In dark mode, component view - info isn't visible
( letters are dark )
In light mode I can see the dark letters
Best

Klaus
 

redmountain

New Member
Thread Starter
Joined
Dec 8, 2022
Posts
31
Opening a file with REWv 5.31.3 with multiple measurements gives funny phase pics
on some measurements...
Not so in REWv5.20.13
 

Attachments

  • Screenshot 2024-12-12 at 09.08.15.pdf
    804.2 KB · Views: 4

redmountain

New Member
Thread Starter
Joined
Dec 8, 2022
Posts
31
Hmmhh...
The impedance measurements are inconsistent here...
Component model available on some measurements and not at all on others
Restarting REW gives another behaviour...
I check with SPL measurements and these are ok
It's impedance that isn't consistent,
at least with my measured ferrite coil mentioned above
??
Don't know how to troubleshot...

Klaus
 

redmountain

New Member
Thread Starter
Joined
Dec 8, 2022
Posts
31
Screenshot and mdat...
 

Attachments

  • Screenshot 2024-12-13 at 12.54.01.pdf
    399.2 KB · Views: 4
  • Impedance measurement component model not available.mdat
    897.9 KB · Views: 3

redmountain

New Member
Thread Starter
Joined
Dec 8, 2022
Posts
31
10 minutes later...
 

Attachments

  • Screenshot 2024-12-13 at 13.25.58.pdf
    420.1 KB · Views: 4
  • Impedance measurement component model available.mdat
    897.9 KB · Views: 1

John Mulcahy

REW Author
Joined
Apr 3, 2017
Posts
8,115
Your measurements become extremely noisy at high frequencies, so much so that the model fit can fail. You can help that by using a lower end frequency for the measurement and by applying some smoothing. You could also start the measurements at a lower frequency, e.g. 10 Hz, to provide some additional data.
 